Registration errors in compound b-scans
阅读量:
7
摘要:
The effect of B-scanner articulated arm angle measurement errors on B-mode registration is considered. For a single-angle articulated arm model, a mathematical equation predicting B-mode misregistration as a function of angular errors and compound scan angle is derived and discussed. This model accurately predicts registration errors for transducer connector/acoustic axis misalignment or third arm angle errors.
展开
关键词:
Calibration B-mode registration Scanning errors Transducer fabrication Connector/acoustic axis alignment
DOI:
10.1002/jcu.1870090107
被引量:
年份:
1981