Randomized Evaluation of an Inflatable Femoral Artery Compression Device After Cardiac Catheterization
摘要:
Mechanical femoral artery compression devices have several limitations. We compared a novel disposable beltheld pneumatic compression device to manual compression alone in 213 patients randomized into two equal groups. Both were comparable for age, gender, current therapy with aspirin (ASA) and warfarin, diameter of the arterial sheath, previous procedures via the same artery, procedure duration, and blood pressure. Manual compression time was 12 ± 3 minutes. Pneumatic compression was reduced during 60 minutes. Patient discomfort was assessed as none (82% vs 88%), mild (13% vs 8%), moderate (3% vs 4%), or severe (2% vs 0%) for the manual versus pneumatic group, respectively. Bleeding and hematoma occurred in 7.5% of patients with no difference between the treatment groups. However, manual compression was significantly more effective in the higher range of systolic blood pressure, and pneumatic in the lower range, with a cut point of approximately 170 mmHg. Predictors for bleedin